DINAMAP SuperSTAT NIBP technology meets ANSI-AAMI standard for neonatal and infant population

Back to Clinical View Home Page

PDF icon Download Resource

This study1 tested the accuracy of the DINAMAPTM SuperSTAT blood pressure technology in neonates and infants according to the ANSI/AAMI SP-10:2002 Standard.

The study took place in neonatal intensive care units and cardiac cath labs. Systolic, diastolic, mean arterial pressure (MAP), and pulse rate (PR) were compared to intra-arterial blood pressure (IBP) measurements. The differences between the SuperSTAT algorithm measurements and the IBP reference were analyzed.

The SuperSTAT NIBP algorithm met the ANSI/AAMI SP-10:2002 accuracy specifications of ±5 mmHg for mean difference and ≤ 8 mmHg standard deviation for the neonatal/infant population.

1 Diego Lozano, MD., Andrea Stebor, RN, Ph.D., Rana F. Mikhail, Ph.D

 