Ultrasound Transducers - Care, Safety and Cleaning
WARNING: Ultrasound transducers are highly sensitive medical instruments requiring proper care and cleaning.
Failure to follow these guidelines and those provided in the Basic Users Manual may result in:
-
Severe lens damage
-
Image quality degradation (not covered under the GE product warranty)
|
|
|
Transducer Care and Maintenance
GE Transducer Care & Safety Guidelines
-
Handle all transducers with care.
-
Prevent damage to transducers by placing them in their holder, carrying case or storage drawer when not in use.
-
Do not drop or subject transducer to any type of mechanical shock. Impact may compromise probe operation, safety features or result in sharp edges that could damage the protective sheath and/ or injure sensitive tissue.
-
Inspect transducer lens, cable & housing before each exam.
-
Do not use damaged transducers. Injury to the operator or patient may occur if cracks, cuts, sharp edges or exposed wiring exist. Cleaning and/ or gel solutions may leak into the transducer resulting in electrical shock. Discontinue use and notify the GE Service representative.
-
Avoid unnecessary stress or bending to cable.
-
Do not twist, kink or pinch cable. Excessive bending or stress on cable may result in damage to its insulating properties causing shock to the patient or operator.
-
Strictly follow the immersion table provided in this guide when cleaning & disinfecting probes.
-
Do not soak transducer longer than recommended by germicide manufacturer.
-
Do not immerse transducer connector in germicide.
-
Use only GE-compatible gels and germicides and strictly follow the manufacturers instructions when applying, cleaning & disinfecting transducers.
-
Do not use non-recommended coupling gels that contain lotions, mineral oil, olive oil, lanolin, polyethylene glycol, dimethylsilicone, methyl or ethyl parabens.
-
Do not clean transducers with non-recommended germicides that contain methanol, ethanol, bleach or alcohol.
-
Do not steam, heat autoclave or use ethylene oxide (EO) gas processes on general surface or TE/TV/TR transducers.
|
GE Transducer Cleaning Guide
|
Note: Prevent disease transmission by properly cleaning & disinfecting transducer surfaces (between patients) that come in contact with body tissue.
-
Refer to the APIC Guidelines for Selection & Use of Disinfectants.
-
Always use sterile, legally marketed transducer sheaths for intra-cavitary, intra-operative and biopsy procedures.
-
GE does not substantiate the claims of infection control capabilities of recommended germicides.
-
Strictly follow the germicide manufacturers instructions for duration of contact time.
-
Failure to do so may result in ineffective decontamination, as well as, severe lens damage & transducer failure not covered under the GE product warranty.
|
Cleaning - All Transducers
-
Disconnect transducer from ultrasound console and rinse probe with a warm nonabrasive soap & water solution.
-
Meticulously scrub transducer as needed with a soft brush, sponge or gauze pad to remove all gel and bio-residues.
-
Air dry or dry with soft cloth or gauze pad.
|
Low Level Disinfecting: General Surface Transducers
|
After cleaning transducer: Spray, wipe or immerse transducer in a disinfectant (see Table) as directed by the germicide manufacturer. Thoroughly rinse with sterile water and pat dry with a soft sponge, cloth or gauze pad.
|
High Level Disinfecting: TE/TV/TR Transducers
|
After cleaning transducer: Immerse transducers that come in contact with mucous membranes, blood or other body fluids in a GE-compatible high level disinfectant (see Table).
Strictly follow the germicide manufacturers instructions. Thoroughly rinse with sterile water and pat dry with a soft sponge, cloth or gauze pad.
|
|
|