MAC 1200

MAC® 1200 Product Specifications

The MAC 1200 Resting ECG System offers the quality and convenience features users appreciate. The system can be configured and customized to meet the needs of hospitals, clinics and office-based practices. In addition, this system allows physicians to affordably access ECG data gathered throughout a provider network, assuring the highest possible standards in quality and continuity of care. Available options include a cart, carrying case and modem.

Performance Specifications

Pre-acquisition: Provides 10 seconds of instantaneous ECG acquistion and printing of any selected report format

Microcomputer-controlled 12-lead resting ECG System

SIGNAL PROCESSING

Signal input: Type CF according to IEC

Digital sampling rate: 1000 samples/second/channel

Dynamic range: Differential signals for AC voltage ± 10mV
Superimposed DC voltage (Polarization voltage) ±600mV

Resolution: 5 µV

Frequency response: 0.04 - 150 Hz

Common mode rejection: > 140 dB

Input impedance: > 100 M

Leakage current: < 10 µA

Pacemaker detection sensitivity: All leads,duration 0.1 - 2.5 ms, amplitude > 5 mV

DISPLAY

Type: LCD 320 x 240 pixel, backlit Contrast adjustable

Displayed data: Selected lead group, operation mode, lead check, heart rate, AC filter, muscle filter, ADS, gain, speed, name, ID, 3 ECG waveforms

WRITER

Type: Thermal array print head

Resolution: Vertically 8 dots/mm, horizontally 25µm at 25 mm/s

Paper: 8.5 x 11 inch

Paper type: Z-fold

Writer speed: 5, 25, 50 mm/s

KEYBOARD

Type: Membrane keypad with tactile feedback / 46 keys

SPECIAL FUNCTIONS

Anti Drift System (ADS): Automatic compensation of baseline fluctuations caused by polarization voltage fluctuations at the electrodes

Baseline adjustment: ECG presentation on averaged baseline

Electrode monitoring: Audible and visual indication on the LCD of disconnected electrodes or line break. Each single electrode is monitored

Copy: In the automatic 12-lead mode, after ECG recording, copies of the ECG can be printed

Test: Automatic performance test upon power up, including verification of the signal path starting at the signal input

POWER SUPPLY

Ratings: 95-240 V (0.2-0.6A)

Frequency: 49-65 Hz

Battery type: NiCad, 18V, 1.3 Ah

Battery charge time: 4 hours

Physical Specifications

Width: 14.5 in (370 mm)

Height: 3.7 in (94 mm)

Depth: 12.6 in (320 mm)

Weight: 12.3 lb (5.6 kg) with battery