GE Medical Systems
Search
My Account
Home > GE Community > MRI> > Ortho MRI Printer Friendly


Back to Orthopedic MRI Home
Chapter 6: The Hand and Wrist
David Stoller, MD

6.2 - Dave Stoller's Hand and Wrist Protocol
6.2.7 - Optional GRE Axial Images
Timing
Parameters
Series 7
This is performed when there is inadequate fat suppression. T2*-weighted images are obtained using a GRE sequence. These show intratendinous signal better than T2-weighted images, but do not depict bone contusions reliably.

 
Axial
T2* 2D-GRE
TR/TE
500/18
ETL
Flip Angle
23
RBW
12.5kHz
FOV
8cm
Matrix
256 x 256
ZIP512
no
Slice Thickness
3mm
Interslice Gap
0.5mm
NEX
2
Acquisition Time
4:20
Fat Sat
no
Tailored RF
no
No Phase Wrap
yes
Flow
Compensation
no
Freq
R/L


T1-Weighted Axial Wrist
Return to Table of Contents