6.2 - Dave Stoller's Hand and Wrist Protocol
6.2.5 - Axial Images
Timing
Parameters
Series 4
Series 5
Axial images are obtained with T1-weighted, and fat-suppressed T2-weighted FSE sequences. If the fat suppression is inadequate in the axial plane, these images may require supplementation with a T2*-weighted GRE technique, either 2D or 3D. Axial images must extend from the distal radioulnar joint through the proximal metacarpals.
Axial
PD FSE-XL
Axial
FSE-XL FS
TR/TE
3000/36
3000/36
ETL
8
8
Flip Angle
RBW
13.89kHz
13.89kHz
FOV
8cm
8cm
Matrix
320 x 256
256 x 224
ZIP512
yes
yes
Slice Thickness
3mm
3mm
Interslice Gap
0.5mm
0.5mm
NEX
2
3
Acquisition Time
3:22
4:37
Fat Sat
no
yes
Tailored RF
yes
yes
No Phase Wrap
yes
yes
Flow
Compensation
no
no
Freq
R/L
R/L


PD-Weighted Axial Wrist

Return to Table of Contents
Advance to Next Section