When a patient is
placed in a strong magnetic field such as is used for MR
imaging, a tiny amount of magnetization from the protons in
his or her body is created. The exquisite sensitivity of the
MR imaging process allows us not only to perceive this
magnetization, but also to manipulate it in ways that give us
information about the tissues themselves. A "net
magnetization" results from the added microscopic
magnetization vectors from all the tissue protons within the
sensitive volume of the receiving RF coil. (There is also
magnetization in the body outside the sensitive volume of the
RF coil, but because this is not received as part of the MR
signal, it is generally ignored). This net magnetization is
aligned by default (i.e. in its "relaxed" state), with the
direction of the polarizing, or main, magnetic field,
B0.
All MR imaging techniques require an
initial "excitation" of this net magnetization, i.e., the
addition of energy to protons in the body in a way that tips
the net magnetization away from the main field toward the
transverse plane. This magnetization has a tendency to realign
with the main field after a time, or "relax" back to its
longitudinal starting state. The excitation of this
magnetization away from the longitudinal axis is accomplished
by using an RF pulse tuned to the resonant frequency of the
protons. This resonant frequency is proportional to the
polarizing field strength; at 1.5 Tesla, the resonant
frequency is approximately 63 MHz. T2 (and T2*) effects relate
to the behavior of this magnetization when it is in the
transverse plane, and T1 describes the rate at which it
relaxes back to its low-energy state, i.e. aligned with the
main magnetic field.
In a simple pulse sequence, the
net magnetization is tipped 90° into the transverse plane,
where it precesses (i.e. rotates) at its resonant frequency
around the direction of B0. This precessing
magnetization will create an electric current, i, in a
properly positioned RF coil that is tuned to the protons'
resonant (or "precession") frequency. (See Fig.1.12) The
resulting electric current is an MR signal that can be
collected and processed to form an image.
|
|
Figure 1.12 The precessing transverse
component of tissue magnetization, M(t), creates a
current, i(t), in the RF coil that is detected as the MR
signal. |
Figure 1.12 shows the
behavior of the net magnetization in time. The vector that
corresponds to the net magnetization is precessing in the
transverse plane at a rate determined by its resonant
frequency. As the magnetization vector continually rotates
through the plane of the RF coil, it interacts with the
electrons in the coil, and causes them to slosh alternately
backward and forward. This is the cause of the current which
goes from positive to negative at the precession frequency of
the magnetization, and is detected as the MR signal*. The
degree to which the precessing magnetization is able to push
the electrons backward and forward in the RF coil and create
this alternating current is referred to as the strength of
"coupling" between the tissue and the coil. Ideally, this
coupling is as large as possible so that the maximum current
is produced in the coil for the amount of magnetization that
is created.
Figure 1.12 and the depiction of the net
magnetization as a single vector require some explanation. The
net magnetization that is depicted in this figure is actually
a vector sum of all the magnetization within the tissue to
which the RF coil is sensitive. Magnetization at different
locations within the tissue can behave differently, depending
on the particular local magnetic environment of the protons at
these locations. The precession frequency of a proton is
determined by its specific local magnetic environment, so is
usually slightly different from the idealized resonant
frequency corresponding to the main field strength. It is
actually this variation in precession frequencies for local
magnetic environments that creates contrast between different
tissues, and is also the foundation of the imaging process
itself. Often, physicists will speak of "ensembles of spins".
This is a notation that refers to a group of protons that has
the same resonant frequency. Throughout this discussion, the
magnetization corresponding to an ensemble of spins will be
referred to as a magnetization component. The magnetization
components can be thought of as smaller magnetization vectors,
each of which corresponds to a group of protons with a common
precession frequency (these add to give the detected net (or
"total") magnetization.
The "driving" effect of the RF
pulse causes the magnetization components to precess at the
same frequency with the same phase during the application of
the pulse.† The net
magnetization from these protons is a maximum immediately
after the RF pulse is applied. After the RF pulse is ended,
each of the individual components begins to revert to its own
individual resonant frequency. The coherence between these
individual components decreases as time goes on (i.e. they
fall farther out of step with each other), and the net
magnetization decreases. The effect on the received MR signal
is a decay of its amplitude over time. The time for the MR
signal to decay to 37% of its initial value (i.e. the time
constant for the decay) is referred to as T2*.
In an
imaging pulse sequence, imaging gradients are used to tag the
magnetization in different voxels with specific resonant
frequencies. These frequency signatures are used to map the MR
signal back to its correct anatomical location in the
reconstructed image. The imaging process can be thought of as
a coarse compartmentalization of the magnetization into
discrete components corresponding to individual voxels.
However, within any tissue voxel, there exists a secondary
spread of resonant frequencies. This spread of resonant
frequencies results from differences in the magnetic fields in
each of the individual protons' environments. The magnetic
field at the location of an individual proton is the result of
adding up the individual magnetic field contributions from
many different sources and is not uniform across the entire
voxel. Changes in any of the individual field components from
location to location within a voxel results in a spread of
resonant frequencies within the same voxel.
The
sources of magnetic field at the level of individual protons
are many; however, a first approximation can be arrived at by
considering only the most important contributions. First, any
main magnetic field imperfections contribute to the loss of
coherence of the transverse magnetization. Second, even in a
perfectly created scanner, the main field intensity would
never be completely uniform on the sub-voxel level in a
clinical imaging situation. Regions of slightly different main
field strength exist throughout each voxel due to differences
in tissue magnetic susceptibility. Magnetic susceptibility is
a parameter that describes the extent to which a substance
will magnetize when placed in a uniform magnetic field. At
boundaries between tissues of different susceptibilities,
magnetic field gradients are created. These gradients cause
protons within the same voxel to precess at different
frequencies, and therefore contribute to a faster T2* decay of
the MR signal. The presence of metal, or an interface with air
also creates spatial variations of field strength due to the
extreme difference of magnetic susceptibility with tissue.
Third, protons on different molecules have different resonant
frequencies due to their "chemical shifts",( i.e. different
local magnetic fields created by different configurations of
neighboring electrons within the molecules themselves). All of
these three contributors to the T2* decay are static in time
(at least for the length of time between the excitation pulse
and the acquisition window), and are referred to specifically
as T2' effects.
In addition to these static T2'
effects, there exist other contributing factors to the
observed T2* decay that arise from more ephemeral
interactions. Within each of the macroscopic magnetic
environments that make up a voxel, there exists another layer
of magnetic interactions that causes the many millions of
protons that may exist there to fall out of sync with each
other. These interactions are known as T2 effects, and occur
on the molecular and submolecular level. Each proton feels the
magnetic fields that are created by neighboring protons and
other magnetic nuclei such as Gadolinium (Gd3+).
The individual resonant frequency of a single proton at any
particular moment depends on that proton's immediate magnetic
environment. This causes the individual protons to precess at
different frequencies and to lose step (phase coherence) with
their neighbors, even though their macroscopic magnetic
environment may be the same. Because the protons are
constantly in motion, the immediate magnetic environment of
each proton changes throughout the excitation-to-acquisition
time. Generally speaking, the more mobile the protons are in a
tissue, the more similar their individual precession
frequencies will be to an average frequency. Tissues with
mobile protons will therefore have longer T2s (i.e. slower T2
decay rates) compared to tissues with bound or restricted
protons (assuming that there is no external contrast agent
like Gd3+ present). Synovial fluid, edema, and necrotic cores
of tumors are examples of tissue components with long T2.
Tissues with Type I collagen such as fibrocartilage (meniscus,
labrum and tendons) tend to have shorter T2s.
* Actually, the MR signal
is detected as a voltage, rather than a current, but these
quantities are related simply by the electrical resistance of
the coil.
† An analogy
for frequency and phase is the lock-steip march of soldiers:
not only do they all march with the same number of steps per
minute (frequency), but they also all raise their right feet
at exactly the same time
(phase).