Ideally, we would like
to obtain our images from contiguous slices. In practice,
there is always some excitation outside the desired slice
boundaries (usually referred to as slice "cross-talk"). This
means that when exciting a particular slice with an RF pulse,
neighboring slices will also experience some partial
excitation at the edges nearest to the excited slice. Each
pulse sequence uses different RF excitation pulses, and the
degree of out-of-slice excitation is therefore particular to
the pulse sequence. In general, if slices are acquired using
an interleaved ordering scheme*, it is possible to use
zero gap without noticing any loss of signal. However, if a
sequential acquisition is used, inserting small gaps
between slices may be necessary. For most pulse sequences, we
recommend using gaps of approximately 10% of the actual slice
thickness. (See Figure 1.10)
|
|
Figure 1.10 Due to the finite length of RF
pulses used in imaging pulse sequences, the slice
profiles of the pulses are imperfect, resulting in some
excitation of the neighbouring slices. An
interleaved acquisition helps to reduce the
amount of signal loss that occurs as a result of
imperfect slice profiles. |
In orthopedic imaging,
a very long TR is often used in order to acquire the large
number of thin slices that are required for the desired
coverage. The default acquisition scheme for FSE is that the
odd slices are acquired first, then the even slices within the
same TR. If the TR is long (~4000ms), enough time will have
passed between excitation of neighboring slices to allow
significant recovery of the out-of-slice magnetization. Zero
gap can usually be used in this case with very little
noticeable signal loss from slice crosstalk.
* An interleaved
acquisition means that the odd slices are acquired in the
first pass (i.e. Slices 1,3,5,7,...) and the even slices are
acquired in the second pass (Slices 2,4,6,8,...)