If the receiving RF
coil is sensitive to tissue signal arising from outside the
desired FOV, this undesired signal may be incorrectly mapped
to a location within the image, a phenomenon known as
aliasing. This is a consequence of the acquired k-space
frequencies not being sampled densely enough, whereby portions
of the object outside of the desired FOV get mapped to an
incorrect location inside the FOV. To prevent this effect in
the frequency-encode direction, data oversampling is used and
is performed automatically on the GE Signa scanner. In this
method, the signal is sampled at a higher bandwidth than
required, and a digital filter is applied to re-sample the
data at the user-specified RBW. The internal high bandwidth
sampling prevents this aliasing phenomenon in the
frequency-encode direction, and filtering to provide the
user-specified RBW improves the SNR.
A similar problem
occurs in the phase-encoding direction, where the phases of
signal-bearing tissues outside of the FOV in the y-direction
are a replication of the phases that are encoded within the
FOV. This signal will be mapped, or "wrapped" back into the
image at incorrect locations, and is known as "phase-wrap"
artifact. No Phase Wrap is a user-selectable parameter
that maps this signal to its correct location outside the FOV,
then discards any signal from outside the FOV before
displaying the image. In the shoulder, for instance, No
Phase Wrap should be used for coronal scan planes to
eliminate signal originating from the chest medial to the
shoulder. When using a shoulder coil, the signal drops off
rapidly beyond the area of interest; however, there is still
enough signal generated from outside the desired FOV to result
in extremely distracting phase wrap artifact and the use of
No Phase Wrap is therefore recommended.
No
Phase Wrap works by filling k-space to the same extent,
using twice as many phase-encoding steps (See Figure 1.7).
(Because twice as many phase-encoding steps are required, the
selected NEX must be two or greater when No Phase
Wrap is selected).
|
|
Figure
1.7 No Phase Wrap works by acquiring twice as
many phase-encode lines in the same extent of k-space.
This results in a doubled FOV in the y-direction with
the same spatial resolution. the two outer halves of the
image which contained the unwanted signal are then
discarded. |
K-space is filled to
the same outer extent, and the spatial resolution in the image
space is therefore unchanged. However, using twice as many
phase-encoding steps means that the FOV in the phase encode
direction is doubled.* Following image reconstruction using an
FFT, the two outer halves of image space are discarded. It
should be noted that a sqrt(2) improvement in SNR occurs for a
2 NEX scan when the No Phase Wrap option is
selected. Figure 1.8 shows an axial knee image with and
without No Phase Wrap selected.
|
|
Figure
1.8 (Image reprinted with permission from: MRI
Principles, by Donald G. Mitchell, W.B. Saunders
Company, 1999). |
* K-space and image
space have the following reciprocal relations: (1) Extent in
k-space determines image spatial resoultion; (2) Resolution in
k-space determines extent in image space (FOV).