1.2.4 - Receive
Bandwidth
An alternative, but
equivalent way of understanding why SNR depends on RBW is to
consider the frequency-domain behavior. Because thermal noise
is spread uniformly across the whole radiofrequency band,
decreasing the RBW means that less noise is collected along
with the MR signal. For a given FOV, the smallest possible RBW
will yield the highest SNR. For a given FOV, it is possible to
acquire an image using a range of RBWs. As illustrated in
Figure 1.4, the combination of the RBW and FOV together
determine what the readout gradient amplitude (Gx) must be to
map this range of frequencies to the specified FOV. On a Signa
scanner, Gx is changed automatically by the system, with the
user selecting both the FOV and RBW.
|
|
Figure
1.4 The RBW and FOV are user-selected on the Signa
scanner, and the gradient strength in the readout
directin, Gx, is adjusted appropriately. For larter RBW,
the gradient strength must be increased to keep the FOV
the same. |
There is a limit to
how small a RBW can be selected for a particular choice of
timing parameters. Decreasing the RBW requires that the signal
must be sampled more slowly. Thus, for the same number of
acquired data points, Xres, the acquisition window is
necessarily longer. In a spin echo, or gradient echo pulse
sequence, the length of the acquisition window is limited by
the userŐs choice of TE. If the acquisition window is too long
to fit in the specified TE, an error will be generated. In a
fast spin echo (FSE) pulse sequence, the choice of RBW is
flexible (when Variable Bandwidth is selected), and is
specified by the user. (See Fig. 1.5) However, the effective
TE for FSE (TEeff ) must be adjusted by the scanner
from the user-selected value, based on the minimum echo
spacing (ESP) that can be achieved for the chosen RBW.
TEeff is defined as the echo that is used to fill the
center of k-space in an FSE acquisition, and must be an
integral multiple of the ESP. Shorter ESP (obtained using
higher RBW) therefore means that the user can more precisely
select the TEeff for the sequence. (For a more detailed
discussion of the FSE sequence, see Section 1.4.3).
|
|
|
Figure 1.5 The RBW specifies the rate at which
data is sampled in the acquisition window. For higher
RBW, the data is sampled more rapidly, and the
acquisition window for a given Xres is shorter.
|
While a lower RBW is preferred for higher
SNR, chemical shift misregistration and
susceptibility-induced distortion artifacts are more
severe. A lower RBW for a constant FOV results in lower
bandwidth per pixel, i.e. each pixel location is encoded
with a narrower range of frequencies. This means that MR
signals at frequencies different from the central
frequency of water will be mapped to pixels farther away
from their actual location, resulting in more pronounced
susceptibility and chemical shift artifacts. As will be
discussed in more detail in Chapter 10, post-surgical
patients who have metal instrumentation should be imaged
at higher RBW to reduce the imaging distortions
resulting from susceptibility gradients created by the
magnetized metal components. For these patients,
reducing image distortions from metal artifact is a much
more important consideration than maximizing the image
SNR. | |
Return to Table of
Contents | Advance to Next
Section | | |