1.2 - Imaging
Parameters: Effects on Image SNR |
1.2.1 -
Measuring Image SNR
|
Ideally,
a homogeneous material would have a uniform appearance on an
MR image, i.e. all pixels corresponding to a uniform material
would have the same grey-scale value. In practice, however,
noise is always superimposed on the images, and is manifested
in the image as a fluctuation of the pixel values around a
central, average value for a uniform material. Noise in an
image is most obviously apparent as a fluctuation of
grey-scale pixel values in the image background where, under
ideal circumstances, the signal intensity would be zero.
|
 |
In a well-designed clinical MRI system, the noise in an image
results mainly from the body itself. Because the human body is
warm, it emits thermal energy at a wide spectrum of
frequencies. This includes the well-known infrared energy that
allows warm bodies to be imaged in the dark using
infrared-sensitive detectors, as well as radiofrequency (RF)
energy at the same frequency as the MR signal itself. This
thermal radiofrequency energy
|
|
is "picked up" by the RF coil
and fed back to the MRI scanner as thermal "noise". Thermal
noise created by the body is thus superimposed on the actual
MR signal, causing an uncertainty in the amplitude of the
measured signal. Noise can also be introduced from the RF coil
as well as other electronic components in the signal reception
chain. These components must be carefully optimized so that
noise arising from sources outside of the body itself is
minimized.
The image SNR is a ratio of the amplitude
of the actual, desired MR signal from tissue to the noise in
the image. Image quality is critically dependent on SNR; the
SNR reflects how much certainty can be associated with the
displayed signal values for the tissue that is being assessed.
In general, it is desirable for the MR signal from tissue to
be large compared to the size of the background signal
fluctuations due to noise. The usual standard for measuring
SNR in an image is the ratio of the mean signal intensity in
the anatomic region of interest (ROI) to the standard
deviation in a region in the background. Different tissues
within the image will have different SNR values due to their
different signal intensities, but the noise in the image is
constant.
Because the thermal energy produced by the
body cannot be reduced, an MRI system must be optimized to
provide the highest possible signal from tissue. The most
critical design component of the imaging system for improving
SNR in an image is the RF coil. The RF coil should be designed
to "couple" strongly to the tissue being imaged, i.e. the
magnetic interaction with the tissue of interest at the MR
signal frequency should be maximized. Note that the magnetic
coupling that occurs between the RF coil and patient will vary
with the patient size and composition. The SNR for a
particular tissue imaged using the same RF coil can therefore
vary from patient to patient due to the different coupling
characteristics of the coil with different patients and
patient positioning relative to the coil. In general, a good
RF coil is designed to provide high signal levels (i.e.
sensitivity) from the tissue of interest over a wide range of
patient sizes, and to create a minimal amount of additional
extra noise from thermal energy of its own electronic
components.
The choice of pulse sequence and imaging
parameters also affects the image SNR, and should be taken
into consideration when designing protocols or assessing an
image for diagnostic information. In the remainder of Section
1.2, the impact on SNR of imaging parameters common to all
pulse sequences is discussed. In MR imaging, it is often
necessary to strike a compromise between two counteracting
image attributes, e.g. higher SNR can be achieved with any
given protocol at the cost of increased acquisition time. It
is important to understand how user-selectable parameters
affect SNR, so that appropriate choices and trade-offs can be
made. |
|
Return to Table of
Contents | Advance to Next
Section |
|
|