Chapter 8: The Knee

8.2.5 - Coronal T2-weighted Images

Timing
Parameters
Series 4
Coronal
T2 FSE-XL FS
TR/TE
4000/72
ETL
8
Flip Angle
RBW
21 kHz
FOV
16 cm
Matrix
512 x 192
Slice Thickness
4mm
Interslice Gap
1mm
NEX
2
Acquisition Time
3:12
Fat Sat
yes
No Phase Wrap
yes
EDR
no
Freq
S/I

The coronal T2-weighted images (Fig. 8.4) are prescribed to match the coronal T1-weighted images. These images are useful for detection of marrow edema, injuries to the collateral ligaments, and cartilage defects.

Figure 8.4 (click on image to see enlarged view)
Coronal T2-weighted image showing fluid at the lateral aspect of the joint (arrow).

Fat suppression is used to improve visualization of fluid signal. Extending the bandwidth to 20-22 kHz allows 512 matrix imaging without an increase in the scan time. Note that the echo time may be reduced to 40-60 msec to improve the SNR of these images.

Return to Table of Contents