|


LightSpeed VCT XT® evaluation system clinical case
VolumeShuttle: CT Perfusion Imaging Made Easy
A new scanning technique for stroke and oncology patients
delivers large coverage, excellent images and faster results
without dose penalty.
By Uri Shreter
|
|
VolumeShuttle™
|
VolumeShuttle™ is a new scanning technique of
LightSpeed® VCT XT that delivers a larger acquisition
coverage than any other CT systems for perfusion
examination. VolumeShuttle™ provides the 80 mm coverage
margin needed, to allow for patient variability in the
Circle of Willis and from the basal ganglia to lateral ventricles,
in one single contrast injection.
CT perfusion (CTP) imaging provides quantitative maps and
measurements of vasculature morphology and function that
have high diagnostic value.
The technique is gaining acceptance for diagnosing ischemic
stroke and determining the appropriate treatment.1.2 It also
shows great potential to help doctors evaluate tumor
angiogenesis and response to treatment3 by detecting
changes in blood flow, blood volume and capillary
permeability.
The new scanning technique and analysis software
developed by GE Healthcare now stand to expand
the diagnostic power of CTP.
Until now, CTP in neurology and oncology has had limited
anatomic coverage on advanced CT scanners. Because the
procedure requires repeated scans at the same location in
short intervals, CTP is traditionally performed with no table
movements between scans, for coverage equal to the width
of the detector, typically 19 mm to 40 mm.
Now, VolumeShuttle™ scanning on the GE LightSpeed® VCT XT
scanner lets clinicians double the anatomic coverage in CTP,
without a dose penalty, while producing high-quality images.
-
80 mm coverage in one single injection
-
Coverage for basal ganglia to top of lateral ventricles
-
Fast table acceleration & deceleration between table locations
|
|
|
VolumeShuttle provides rapid scanning at two table positions
during one bolus injection of contrast.This enables clinicians
to follow, measure and accurately analyze the contrast wash
in and out of blood vessels and tissue. The technique takes
advantage of key attributes of the LightSpeed VCT XT scanner:
-
A 40-mm-wide detector array with 64 rows of
0.625-mm elements
-
Advanced systems for controlling X-ray generation and
patient table movements.
-
Advanced image-reconstruction algorithms.
-
Innovative analysis algorithms and software.
|
Clinical need
|
Conventional CTP scanning provides blood flow and
morphology information in a slab of anatomy using repetitive
scanning at 0.5- or 1.0-second intervals for 30 to 60 seconds.
The data is then analyzed to generate detailed maps of
regional blood flow (BF), blood volume (BV) and mean transit
time (MTT). Figure 1 shows perfusion maps in a stroke
patient.
The thickness of the scanned slab in this technique is limited
to the width of the detector. The concern is that important
information, such as a brain infarct or a highly perfused
component of a tumor, may lie outside the scanned volume
and so cannot be seen.
One response to the coverage limitation is to perform two
consecutive series of scans at contiguous locations, each
with its own contrast injection. However, this doubles the
contrast load, doubles the effective radiation dose, and
lengthens the exam, potentially delaying sometimes-critical
treatment decisions.
The VolumeShuttle mode instead scans alternately at two
contiguous locations in one exam and with one contrast
injection. During an exam, scans can be performed at
3.0-second intervals for 60 seconds – a total of 20 scans.
This technique provides all the information needed to
calculate perfusion maps and parameters, while doubling
the volume covered in conventional CTP scans, reducing
the radiation dose, and preserving clinical accuracy.
|
Advanced technology
|
To be clinically useful and accurate, VolumeShuttle scanning
has two key requirements:
1. Advanced control of table movement and X-ray beam timing.
The table must move rapidly back and forth between the
two scan locations with minimal delays in positioning and
in triggering of the X-ray exposure. This ensures enough
sampling frequency at each location to enable extraction
of blood-flow parameters.
2. Precisely repeatable table position. The table must stop
exactly at the same two scanning locations. Otherwise,
there will be misregistration of pixels over the course of the
exam, and the perfusion maps will be inaccurate –
just as they would be if the patient moved.
The LightSpeed VCT XT scanner precisely controls both the
X-ray beam and the table. As soon as the X-rays are turned
off in one axial scan, the table moves rapidly to the next
position. The patient is then in position for the X-rays to be
turned on again. The scanner’s cradle drive uses a highly
accurate stepper motor drive with two belt reduction
mechanisms. It moves the patient 40 mm (1.6 in.) in
0.7 second – smoothly and quietly.
The patient experiences a maximum acceleration of less
than 1/30 of gravitational pull (G) and a top speed of less
than 0.25 mph (0.4 km/hr), causing negligible discomfort.
The table position is repeatable to within 0.25 mm (0.1 in.)
under full load of 500 pounds.
The scanner also has a precise, high-speed on-off switch for
the X-ray beam, synchronized to table movement and location.
It ensures that X-rays are turned on as soon as needed and
completely off before the table starts moving – thus minimizing
the delay between scans and the radiation dose.
As an example, a typical VolumeShuttle exam with a scan
interval of 3 seconds at each location may deliver only
two-thirds the effective dose of a single-location conventional
scan at the same X-ray kVp and mA. This means a 33 percent
dose reduction for the same image quality and double
the coverage.
|
Proven performance
|
In development of VolumeShuttle scanning, one concern was
that the longer 3.0-second sampling intervals at each scan
location could affect the accuracy of the perfusion maps
and the perfusion parameters calculated from them. To
verify the adequacy of VolumeShuttle scanning for clinical
needs, GE Healthcare collaborated on a study with
a clinical partner.4
The study used data gathered from previous CTP exams
of acute stroke patients scanned within seven hours of
symptom onset. These exams were performed conventionally
with a scan interval of 0.5 seconds, ensuring highly accurate
perfusion maps.
Images from these studies were progressively removed to
simulate a CTP exam with scans performed at up to 3.0-second
intervals. GE CT Perfusion software was used to analyze the
original data and the reformatted data, and the resulting
perfusion maps were compared.
Figure 2 shows a comparison of the perfusion maps, and
Figure 3 shows the good correlation for an exam involving an
interval of 3 seconds between scans. The study concluded
that scan intervals of up to 3 seconds can be used in CTP
scanning while retaining the accuracy of blood flow, blood
volume, and mean transit time values – without dose penalty.
A separate study of brain perfusion scanning, conducted
independently, verified that the VolumeShuttle technique
increased coverage without dose penalty and required only
a single contrast injection without loss of visual diagnostic
accuracy.5 In particular:
-
When compared to conventional CT perfusion with one
bolus injection, the VolumeShuttle mode enhances
diagnostic confidence by covering a larger volume.
-
When compared to two-location, two-injection scanning,
VolumeShuttle reduces overall scan time, providing all
the images needed for diagnosis and treatment.
|
|
Summary
|
CTP scanning in the VolumeShuttle mode takes advantage
of the latest technology on the GE LightSpeed VCT XT scanner
to double the anatomical coverage with one bolus of
contrast and without dose penalty. Clinical studies show that
in terms of image quality and accuracy of quantitative
measurements, VolumeShuttle is an acceptable
alternative to conventional scanning.
|
|
|