LightSpeed™ VCT

SnapShot Pulse: A Revolution in Cardiac Imaging

A new technique for coronary CT angiography delivers excellent image quality at dramatically reduced radiation dose

By Uri Shreter

Technology solution

Until recently, step-and-shoot techniques were not feasible, mainly because limited detector coverage required multiple consecutive axial scans to cover the entire heart, and the delay between scans made the overall scan too long for breath-hold contrast-enhanced imaging. Such techniques also require:

  • Precise, high-speed on-off switching of the X-ray beam, synchronized to table movement and location.
  • Precise synchronization of the X-ray to the ECG signal to ensure both maximum dose reduction and continuity of the acquired data between steps.
  • Advanced algorithms to reconstruct high-quality, thin-slice axial images.


The LightSpeed VCT XT platform combines all these elements to enable a robust prospectively gated axial CCTA scan. Radiation is reduced to less than 5 mSv for medium-sized patients – 70 percent less than for helical scans – with no reduction in image quality.


Acquiring the data

Figure 1 - Illustration of SnapShot Pulse acquisition

Figure 1 - Illustration of SnapShot Pulse acquisition

The main requirement for the SnapShot Pulse technique is that data acquisition occurs at the right heart phase to guarantee high-quality arterial imaging. The system decides when to turn the X-rays on in a given heartbeat before the ECG signal of the next R-wave appears. Therefore, it is critical to predict the R-wave’s timing accurately and to have enough extra data to accommodate variations in heart rate.

SnapShot Pulse accomplishes this by monitoring the ECG signal in real time, analyzing heart rate changes, and adapting the timing of scans accordingly. The technique uses a statistical model based on the length and variability of previous heart cycles to time the next axial scan.

The overall scan time must be as short as possible: The contrast bolus must be caught at the time of maximum and steady enhancement to ensure high visibility of small vessels and to minimize density changes from one axial scan to the next. Short scans also help reduce variations in heart rate during a study.4

The LightSpeed VCT XT scanner is ideal for the SnapShot Pulse mode because its detector covers a wide 40-mm slab in one rotation. This means only three or four such slabs are needed to cover the length of a typical heart (see Figure 1). Ideally, one slab is completed in every heartbeat, and that is possible at low heart rates. At higher rates, it is necessary to skip one beat between slab acquisitions, and the overall exam lasts five or seven heartbeats.

SnapShot Pulse scanning is enabled by fast and precise movements of the table. The scanner uses an advanced system to control both the X-ray beam and the table. As soon as the X-rays are turned off in one axial scan, the table moves rapidly yet smoothly to the next position. The patient is then in position for the X-rays to be turned on again at the right time in the ECG signal.

High-quality image reconstruction is essential for SnapShot Pulse scanning. CCTA requires very thin slices and relatively high current to achieve the best diagnostic results. For the LightSpeed VCT XT scanner, individual 0.625-mm images must be reconstructed from the 64 detector rows in the 40-mm-wide detector array. GE-exclusive algorithms enable accurate 3D volume reconstructions and ensure uniform performance across the detector array.

The method has been validated and documented in scientific research and publications.5


Cardiac CT has seen substantial progress in recent years. Faster scans, better resolution, increased X-ray power and advances in reconstruction and filtration algorithms add up to improved image quality and diagnostic confidence. As a result, cardiac CT has become widely accepted as a highly sensitive tool to help doctors to detect cardiac disease and accurately quantify coronary artery obstructions.

One persistent concern in CT imaging has been the high radiation dose burden in coronary CT angiography (CCTA). Now, a new scanning technique developed by GE Healthcare has addressed that problem. The SnapShot™ Pulse scanning mode, used on the GE LightSpeed® VCT XT scanner reduces radiation dosage as much as 70 percent when compared to conventional helical techniques, while producing highquality images.

In essence, the technique captures a complete picture of the heart using a series of three to four X-ray axial “snapshots” taken at precise patient table positions and timed to correspond to a specific phase of the cardiac cycle. The technique takes advantage of five key attributes of the LightSpeed VCT XT scanner:

  • A 40-mm-wide detector array with 64 rows of 0.625-mm elements.
  • Advanced systems for controlling X-ray generation.
  • Precise patient table movements.
  • Real-time ECG signal monitoring.
  • Advanced reconstruction algorithms.


Meeting a need

Figure 2 - Retrospective gating helical vs SnapShot Pulse acquisition

For years, clinicians and CT scanner manufacturers have agreed on the need for new protocols to reduce dosage in CCTA without sacrificing image quality and diagnostic value. The main reason for the high dose in CCTA is that conventional helical scan protocols use very low pitch factors to ensure continuous data availability for reconstruction of images at each location along the heart at the required heart phase.

In the process, the X-ray beam irradiates each location along the scan path three to five times, even though, in most cases, a large portion of the information acquired during the scan is not used. In reality, as documented by two studies using the GE LightSpeed VCT scanner, CCTA exams using reconstruction from a single heart phase provide the needed diagnostic images.1,2

The logical solution is to confine X-ray exposure as much as possible to a single heart phase, so that the necessary data is captured while the overall X-ray on-time is shortened, and total radiation exposure is thus reduced.

Several variations on this approach have been tried. One of the most successful is a retrospective gated helical acquisition technique called ECG-driven X-ray beam current modulation. This method, available for most high-end CT scanners, uses high tube current for a small range of heart phases to ensure low noise and high image quality during those phases, while reducing current to about 20 percent of the maximum during the remaining phases. Dose is reduced by 30 to 50 percent3.

The GE SnapShot Pulse technique is a natural extension of this method. It uses prospectively triggered axial step-andshoot scans in which X-rays are turned on only during the required heart phase and turned off completely at all other times (see Figure 2). Figure 2 - Retrospective gating helical vs SnapShot Pulse acquisition


Testing performance

Testing performance

Figure 3 - SnapShot Pulse performance chart

GE conducted laboratory tests using various phantoms to measure image-quality parameters including slice sensitivity profile, in-plane resolution, noise, low-contrast resolution, and image uniformity with simulated heart rates from 40 to 65 bpm.6 The results were then compared to those of helical coronary CCTA exams. Quality parameters with SnapShot Pulse compared favorably, while the X-ray dose with SnapShot Pulse was markedly lower (see Figure 3).


Figure 4 - Exam Repartition vs BMI and average corresponding dose

Figure 4 - Exam Repartition vs BMI and average corresponding dose

After laboratory verification that SnapShot Pulse scanning could achieve dose savings without sacrificing image quality, the method underwent clinical testing. In an initial study, 31 patients were scanned with both SnapShot Pulse and conventional retrospectively gated helical scans.7 Patients’ heart rates ranged from 38 to 67 bpm (average 54 bpm), and heart rate variations ranged from 0 to 9 bpm (average 3 bpm). Now, in a recent study, 600 patients were scanned with both SnapShot Pulse and conventional retrospectively gated helical scans with 5 mSv average (Figure 4).


Stats from CCN, Saint Denis, France.

  • 600 patients
  • 5 mSv average
  • Low dose acquisition whatever patient'morphology


Figure 5 - Helical scan images on left, SnapShot Pulse images on the right

Untitled Document


About half the patients required scanning at three locations and five heartbeats to cover the entire heart, and the remainder required four locations and seven heartbeats. Total scan times ranged from 4.5 to 10 seconds (average 6.7 seconds). Total X-ray on-time ranged from 1.2 to 2.5 seconds.

All patients scanned with SnapShot Pulse had diagnostic images of all 15 American Heart Association recommended coronary artery segments. The expert radiologist reviewing the cases rated image quality as excellent and observed identical pathologies in all studies using the two techniques. No pathology shown in the conventional scans was missed in the SnapShot Pulse mode. X-ray exposure with SnapShot Pulse was up to 55 percent lower than in retrospective gating using ECG-driven tube current modulation.

Figure 5 shows representative images from this study with helical scan images on the left and SnapShot Pulse images on the right.

Another study involved 33 patients scheduled for CCTA exams for various clinical indications. All scans, except one with patient movement, yielded clinically acceptable images that the interpreting radiologist rated between good and excellent.

In 12 patients with body mass index (BMI) below 25, scanned for coronary artery evaluation, the average effective dose was 4.6 mSv*. In 16 patients with BMI between 25 and 40, scanned for coronary artery evaluation, the average effective dose was 6.4 mSv*.

The dose for all coronary exams ranged from 1.1 to 7.5 mSv* (average 5.5 mSv*). These doses represent reductions of 50 to 65 percent compared to conventional helical cardiac scanning with ECG-based tube current modulation. Five patients with BMI between 22.5 and 29.7 were scanned for coronary bypass evaluation. The average scanned length for these cases was 270 mm. The average effective dose ranged from 10 to 13.5 mSv* (average 12.5 mSv*), a more than 50 percent dose reduction. Figure 6 shows representative images from this study.


Conclusion

Figure 6 - Representative images from study

Untitled Document

SnapShot Pulse cardiac scanning takes advantage of the latest technologies and algorithms on the GE LightSpeed VCT XT scanner to achieve significant dose reductions in CCTA imaging with image quality as good as or better than that of conventional helical scans. Clinical studies show that SnapShot Pulse is an acceptable alternative to helical scans when only one heart phase is required, which reduces radiation exposure to 1 to 5 mSv, equivalent to only four to twenty months of exposure to average natural background radiation.7