Technology solution
|
Until recently, step-and-shoot techniques were not feasible,
mainly because limited detector coverage required multiple
consecutive axial scans to cover the entire heart, and the
delay between scans made the overall scan too long for
breath-hold contrast-enhanced imaging. Such techniques
also require:
-
Precise, high-speed on-off switching of the X-ray beam,
synchronized to table movement and location.
-
Precise synchronization of the X-ray to the ECG signal
to ensure both maximum dose reduction and continuity
of the acquired data between steps.
-
Advanced algorithms to reconstruct high-quality, thin-slice
axial images.
|
|
The LightSpeed VCT XT platform combines all these elements
to enable a robust prospectively gated axial CCTA scan.
Radiation is reduced to less than 5 mSv for medium-sized
patients – 70 percent less than for helical scans – with no
reduction in image quality.
|
Acquiring the data
|
The main requirement for the SnapShot Pulse technique is
that data acquisition occurs at the right heart phase to
guarantee high-quality arterial imaging. The system decides
when to turn the X-rays on in a given heartbeat before the
ECG signal of the next R-wave appears. Therefore, it is critical
to predict the R-wave’s timing accurately and to have
enough extra data to accommodate variations in heart rate.
SnapShot Pulse accomplishes this by monitoring the ECG
signal in real time, analyzing heart rate changes, and adapting
the timing of scans accordingly. The technique uses a statistical
model based on the length and variability of previous heart
cycles to time the next axial scan.
The overall scan time must be as short as possible: The
contrast bolus must be caught at the time of maximum and
steady enhancement to ensure high visibility of small vessels
and to minimize density changes from one axial scan to the
next. Short scans also help reduce variations in heart rate
during a study.4
The LightSpeed VCT XT scanner is ideal for the SnapShot
Pulse mode because its detector covers a wide 40-mm slab
in one rotation. This means only three or four such slabs are
needed to cover the length of a typical heart (see Figure 1).
Ideally, one slab is completed in every heartbeat, and that
is possible at low heart rates. At higher rates, it is necessary
to skip one beat between slab acquisitions, and the overall
exam lasts five or seven heartbeats.
SnapShot Pulse scanning is enabled by fast and precise
movements of the table. The scanner uses an advanced
system to control both the X-ray beam and the table. As
soon as the X-rays are turned off in one axial scan, the table
moves rapidly yet smoothly to the next position. The patient
is then in position for the X-rays to be turned on again at the
right time in the ECG signal.
High-quality image reconstruction is essential for SnapShot
Pulse scanning. CCTA requires very thin slices and relatively
high current to achieve the best diagnostic results. For the
LightSpeed VCT XT scanner, individual 0.625-mm images
must be reconstructed from the 64 detector rows in the
40-mm-wide detector array. GE-exclusive algorithms enable
accurate 3D volume reconstructions and ensure uniform
performance across the detector array.
The method has been validated and documented in scientific
research and publications.5
|
|
Cardiac CT has seen substantial progress in recent years.
Faster scans, better resolution, increased X-ray power and
advances in reconstruction and filtration algorithms add up
to improved image quality and diagnostic confidence. As a
result, cardiac CT has become widely accepted as a highly
sensitive tool to help doctors to detect cardiac disease and
accurately quantify coronary artery obstructions.
One persistent concern in CT imaging has been the high
radiation dose burden in coronary CT angiography (CCTA).
Now, a new scanning technique developed by GE Healthcare
has addressed that problem. The SnapShot™ Pulse scanning
mode, used on the GE LightSpeed® VCT XT scanner reduces
radiation dosage as much as 70 percent when compared
to conventional helical techniques, while producing highquality
images.
In essence, the technique captures a complete picture of the
heart using a series of three to four X-ray axial “snapshots”
taken at precise patient table positions and timed to correspond
to a specific phase of the cardiac cycle. The technique takes
advantage of five key attributes of the LightSpeed VCT XT
scanner:
-
A 40-mm-wide detector array with 64 rows of
0.625-mm elements.
-
Advanced systems for controlling X-ray generation.
-
Precise patient table movements.
-
Real-time ECG signal monitoring.
-
Advanced reconstruction algorithms.
|
Meeting a need
|
For years, clinicians and CT scanner manufacturers have
agreed on the need for new protocols to reduce dosage in
CCTA without sacrificing image quality and diagnostic value.
The main reason for the high dose in CCTA is that conventional
helical scan protocols use very low pitch factors to ensure
continuous data availability for reconstruction of images at
each location along the heart at the required heart phase.
In the process, the X-ray beam irradiates each location along
the scan path three to five times, even though, in most cases,
a large portion of the information acquired during the scan is
not used. In reality, as documented by two studies using the
GE LightSpeed VCT scanner, CCTA exams using reconstruction
from a single heart phase provide the needed diagnostic
images.1,2
The logical solution is to confine X-ray exposure as much as
possible to a single heart phase, so that the necessary data
is captured while the overall X-ray on-time is shortened, and
total radiation exposure is thus reduced.
Several variations on this approach have been tried. One
of the most successful is a retrospective gated helical
acquisition technique called ECG-driven X-ray beam current
modulation. This method, available for most high-end CT
scanners, uses high tube current for a small range of heart
phases to ensure low noise and high image quality during
those phases, while reducing current to about 20 percent of
the maximum during the remaining phases. Dose is reduced
by 30 to 50 percent3.
The GE SnapShot Pulse technique is a natural extension of
this method. It uses prospectively triggered axial step-andshoot
scans in which X-rays are turned on only during the
required heart phase and turned off completely at all other
times (see Figure 2).
Figure 2 - Retrospective gating helical vs SnapShot Pulse acquisition
|
Testing performance
|
GE conducted laboratory tests using various phantoms to
measure image-quality parameters including slice sensitivity
profile, in-plane resolution, noise, low-contrast resolution,
and image uniformity with simulated heart rates from 40 to
65 bpm.6 The results were then compared to those of helical
coronary CCTA exams. Quality parameters with SnapShot
Pulse compared favorably, while the X-ray dose with
SnapShot Pulse was markedly lower (see Figure 3).
|
|
After laboratory verification that SnapShot Pulse scanning
could achieve dose savings without sacrificing image quality,
the method underwent clinical testing. In an initial study,
31 patients were scanned with both SnapShot Pulse and
conventional retrospectively gated helical scans.7 Patients’
heart rates ranged from 38 to 67 bpm (average 54 bpm), and
heart rate variations ranged from 0 to 9 bpm (average 3 bpm).
Now, in a recent study, 600 patients were scanned with both
SnapShot Pulse and conventional retrospectively gated helical
scans with 5 mSv average (Figure 4).
Stats from CCN, Saint Denis, France.
-
600 patients
-
5 mSv average
-
Low dose acquisition whatever patient'morphology
|
|
Figure 5 - Helical scan images on left, SnapShot Pulse images on the right
Untitled Document
About half the patients required scanning at three locations
and five heartbeats to cover the entire heart, and the
remainder required four locations and seven heartbeats.
Total scan times ranged from 4.5 to 10 seconds (average
6.7 seconds). Total X-ray on-time ranged from 1.2 to
2.5 seconds.
All patients scanned with SnapShot Pulse had diagnostic
images of all 15 American Heart Association recommended
coronary artery segments. The expert radiologist reviewing
the cases rated image quality as excellent and observed
identical pathologies in all studies using the two techniques.
No pathology shown in the conventional scans was missed
in the SnapShot Pulse mode. X-ray exposure with SnapShot
Pulse was up to 55 percent lower than in retrospective
gating using ECG-driven tube current modulation.
Figure 5 shows representative images from this study with
helical scan images on the left and SnapShot Pulse images
on the right.
Another study involved 33 patients scheduled for CCTA
exams for various clinical indications. All scans, except
one with patient movement, yielded clinically acceptable
images that the interpreting radiologist rated between good
and excellent.
In 12 patients with body mass index (BMI) below 25, scanned
for coronary artery evaluation, the average effective dose
was 4.6 mSv*. In 16 patients with BMI between 25 and 40,
scanned for coronary artery evaluation, the average effective
dose was 6.4 mSv*.
The dose for all coronary exams ranged from 1.1 to 7.5 mSv*
(average 5.5 mSv*). These doses represent reductions of 50
to 65 percent compared to conventional helical cardiac
scanning with ECG-based tube current modulation.
Five patients with BMI between 22.5 and 29.7 were scanned
for coronary bypass evaluation. The average scanned length
for these cases was 270 mm. The average effective dose
ranged from 10 to 13.5 mSv* (average 12.5 mSv*), a more
than 50 percent dose reduction. Figure 6 shows representative
images from this study.
|
Conclusion
Figure 6 - Representative images from study
Untitled Document
SnapShot Pulse cardiac scanning takes advantage of the latest
technologies and algorithms on the GE LightSpeed VCT XT
scanner to achieve significant dose reductions in CCTA
imaging with image quality as good as or better than that
of conventional helical scans. Clinical studies show that
SnapShot Pulse is an acceptable alternative to helical scans
when only one heart phase is required, which reduces
radiation exposure to 1 to 5 mSv, equivalent to only four to
twenty months of exposure to average natural background
radiation.7
|
|